TY - JOUR
T1 - Mercury determination at trace levels using membrane preconcentration and benchtop total reflection X-ray fluorescence analysis
AU - Marguí, E.
AU - Queralt, I.
AU - Guerra, M.
AU - Kallithrakas-Kontos, N.
N1 - This work was supported by the Spanish National Research Program (Project References: CGL2013-48802-C3-2-R and CGL2016-78783-C2-2-R). I. Queralt acknowledges the grant for a sabbatical stay from Spanish Ministry of Education (Ref. PRX16/00159). M. Guerra also acknowledges the support of the FCT, under Contract No. SFRH/BPD/92455/2013.
PY - 2018/11/1
Y1 - 2018/11/1
N2 - At present, there is a considerable interest in Hg monitoring in water samples due to its widespread occurrence and the high toxicity of most of its compounds. Hg determination in water samples by means of total reflection X-ray fluorescence spectrometry (TXRF) entails some difficulties due to the high vapour pressure and low boiling point of this element that produce evaporation and loss of Hg from the surface of the reflector during the drying process, commonly employed in sample preparation in TXRF analysis. In a former work we demonstrated the usefulness of PVC-based membranes functionalized with dithizone for Hg preconcentration. Limits of detection as low as 0.8 μg·L−1 Hg were obtained using a 1 kW TXRF system. The main goal of the present research was to improve the extraction system and to apply it in combination with low power TXRF instrumentation (50 W). In this sense, some extraction parameters as well as measurement conditions such as the volume of membrane deposited on the quartz reflector, the agitation speed during extraction, the aqueous sample volume and the measurement time were carefully evaluated to obtain the maximum sensitivity for Hg determination. Using the best analytical conditions, a limit of detection for Hg of 0.3 μg·L−1 was obtained which is adequate according to the limits of current legislation on the quality of water for human consumption. Additionally, a morphological study of the preconcentrated sample on the quartz reflector was performed by optical microscopy and information regarding Hg distribution along the membrane was obtained by μ-XRF analysis. It was found that Hg is not homogenously accumulated on the membrane and thus, the size of the membrane on the reflector is a critical parameter that has to be controlled to ensure the complete exposure of the loaded membrane to the X-ray beam. This fact is of special relevance when using this preconcentration approach in combination with micro-TXRF. Finally, to further prove applicability of the TXRF method for real water sample analysis, different types of spiked water samples were analysed.
AB - At present, there is a considerable interest in Hg monitoring in water samples due to its widespread occurrence and the high toxicity of most of its compounds. Hg determination in water samples by means of total reflection X-ray fluorescence spectrometry (TXRF) entails some difficulties due to the high vapour pressure and low boiling point of this element that produce evaporation and loss of Hg from the surface of the reflector during the drying process, commonly employed in sample preparation in TXRF analysis. In a former work we demonstrated the usefulness of PVC-based membranes functionalized with dithizone for Hg preconcentration. Limits of detection as low as 0.8 μg·L−1 Hg were obtained using a 1 kW TXRF system. The main goal of the present research was to improve the extraction system and to apply it in combination with low power TXRF instrumentation (50 W). In this sense, some extraction parameters as well as measurement conditions such as the volume of membrane deposited on the quartz reflector, the agitation speed during extraction, the aqueous sample volume and the measurement time were carefully evaluated to obtain the maximum sensitivity for Hg determination. Using the best analytical conditions, a limit of detection for Hg of 0.3 μg·L−1 was obtained which is adequate according to the limits of current legislation on the quality of water for human consumption. Additionally, a morphological study of the preconcentrated sample on the quartz reflector was performed by optical microscopy and information regarding Hg distribution along the membrane was obtained by μ-XRF analysis. It was found that Hg is not homogenously accumulated on the membrane and thus, the size of the membrane on the reflector is a critical parameter that has to be controlled to ensure the complete exposure of the loaded membrane to the X-ray beam. This fact is of special relevance when using this preconcentration approach in combination with micro-TXRF. Finally, to further prove applicability of the TXRF method for real water sample analysis, different types of spiked water samples were analysed.
KW - Membranes
KW - Mercury
KW - Preconcentration
KW - TXRF
KW - Water samples
UR - http://www.scopus.com/inward/record.url?scp=85050270969&partnerID=8YFLogxK
U2 - 10.1016/j.sab.2018.07.020
DO - 10.1016/j.sab.2018.07.020
M3 - Conference article
AN - SCOPUS:85050270969
SN - 0584-8547
VL - 149
SP - 84
EP - 90
JO - Spectrochimica Acta Part B: Atomic Spectroscopy
JF - Spectrochimica Acta Part B: Atomic Spectroscopy
T2 - 17th International Conference on Total Reflection X-Ray Fluorescence Analysis and Related Methods (TXRF)
Y2 - 19 September 2017 through 22 September 2017
ER -