Abstract
BACKGROUND: Impedance cardiography (ICG) is a noninvasive hemodynamic monitoring tool which can define hypertensive patients' hemodynamic profiles and help to tailor antihypertensive therapy. This study assesses the concordance between ICG-derived indexes used to evaluate left ventricular performance and transthoracic echocardiography (TTE) in hypertensive patients.
METHODS: In this IMPEDDANS post-hoc analysis, the ICG-derived indexes are compared with TTE by Bland-Altman method. Statistical significance of the relationship between the values obtained was assessed by generalized linear mixed-effects models.
RESULTS: In supine position, Bland-Altman analysis showed good concordance for cardiac output (CO) (mean difference of 0.006 mL/min [-0.120; 0.133]), cardiac index (CI) (mean difference of 0.016 mL/min/m2 [-0.471; 0.504]), pre-ejection period (PEP) (mean difference of -0.216 ms [-4.510; 4.077]), left ventricular ejection time (LVET) (mean difference of -0.140 ms [-6.573; 6.293]), and systolic time ratio (STR) (mean difference of -0.00004 [-0.008; 0.008]). In orthostatic position, good concordance was found for CO (mean difference 0.028 mL/min [-2.036; 1.980]), CI (mean difference -0.012 mL/min/m2 [-1.063; 1.039]), and STR (mean difference -0.101 [0.296; 0.094]). No significant difference between methods was identified by the linear mixed-effects models.
CONCLUSION: The ICG-derived indexes CO, CI, PEP, LVET, and STR in supine position have good agreement with TTE. Therefore, ICG can be used to accurately evaluate left ventricular performance.
Original language | English |
---|---|
Pages (from-to) | 461-467 |
Number of pages | 7 |
Journal | Clinical and experimental hypertension (New York, N.Y. : 1993) |
Volume | 40 |
Issue number | 5 |
Early online date | 27 Nov 2017 |
DOIs | |
Publication status | Published - 2018 |
Keywords
- Echocardiography
- hypertension
- impedance cardiography
- left ventricular performance
- systolic time intervals